|
|
|
1. Turn on the Spec 20 and allow it to warm up for 5 - 10 minutes (left front knob). Set wavelength using the dial on top of the Spec 20. |
![]() |
2. Prepare a BLANK cuvette by adding all solvents EXCEPT the substance to be measured. (Refer to your experimental protocol).
- A BLANK is used to calibrate the Spec 20 so that any absorbance attributable to the solvent and/or glass cuvette can be compensated. By zeroing the Spec 20 to the blank, you will measure only the absorbance due to the substance in question.
3. With no tube in the holder, adjust the meter needle to read infinite absorbance (= 0% transmittance) using the left front knob (= power switch).
4. Using a Kimwipe, wipe off/polish the outside of the BLANK cuvette to remove greasy finger smudges etc. (You might want to wear gloves). Using a wax pencil or Sharpie, make a small vertical mark at the top of each cuvette for alignment in the sample holder.
5. Raise the sample holder trapdoor and insert the cuvette such that the line on the cuvette lines up with the line on the sample holder. Close the lid.
6. Using the right front knob, adjust the meter needle to read absorbance = 0.0 (= 100 % transmittance). This step is called setting the "full scale".
The spectrophotometer is now calibrated to this BLANK. If your experiment involves multiple reaction tube formulations, each one will need its own blank and the Spec 20 must be rezeroed for each.
7. Remove BLANK and insert cuvette containing your sample. Close lid.
8. Read the absorbance (lower scale) OR transmittance (upper scale) as appropriate for your sample.
9. Repeat for subsequent samples which use the same BLANK. (SEE NOTE BELOW)
NOTE: When taking several measurements at the same wavelength over a short time period, you do not need to reblank for each. Over longer times, however, the unit may drift and recalibration to the BLANK will be necessary. IF, however, you change the wavelength, you must re-zero the instrument. If you are taking readings over an extended period or sharing the instrument, re-zero for each measurement.
The Beer-Lambert law describes an important relationship that exists between absorbance (A) and two sample parameters - solute concentration (c) and length of the light path (l). Simply put, the law states that absorbance, A, is directly proportional to c and l, and is represented by the following equation:
In biological research, concentration (c) is usually expressed on a mass/volume basis, e.g., ug/ml, length of light path (l) in centimeters (usually l=1cm), and µ, the absorbance coefficient, is also expressed on the basis of weight. For practical purposes, the light path is the interior diameter of the cuvette and is the same for all samples. Therefore, a plot of absorbance vs concentration yields a straight line with slope µ. Such a curve using known concentrations of a pure substance is called a standard curve. A standard curve is then useful for determining the concentration of the same substance in solutions of unknown concentration. By algebraic rearrangement of the above equation to,
it is clear that concentration can be determined from absorbance
alone. Absorption coefficients for biological molecules can be
determined experimentally or can be found in the literature.
![]()
Modified 9-30-02 gja
Department of Biology, Bates College, Lewiston, ME 04240